The role of deep-learning T2 PROPELLER, diffusion and perfusion sequences in 3.0T MR evaluation of head-neck tumors

By Simona Marzi, PhD, medical physicist, Michele Farella, technologist, Giovanni Di Giulio, technologist, and Antonello Vidiri, MD, Medical Director, Regina Elena National Cancer Institute, Rome, Italy

In evaluating head and neck (HN) cancer, especially in the oral cavity, oropharynx and larynx, obtaining high-resolution images and minimizing motion artifacts is crucial. This is vital for assessing submucosal spread, which is very important for staging and accurate surgical planning. To achieve optimal visualization of tumor-muscle-fat relationships, it is essential to use a 3.0T MR system with T2-weighted sequences. However, motion artifacts may limit the use of 3.0T in these patients. To address this issue, it is generally recommended to acquire a T2-weighted PROPELLER (Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction) sequence in addition to the classic Fast Spin Echo (FSE) T2-weighted sequence. The T2 PROPELLER sequence helps reduce the effects of voluntary and physiological patient motion and magnetic susceptibility artifacts, although at the expense of lower contrast resolution.

 

At our Institute, we utilize SIGNA™ Premier 3.0T with AIR™ Recon DL, a pioneering deep-learning (DL)-based reconstruction software designed to improve SNR and increase image sharpness.

AIR Recon DL is not a post-processing technique, rather it is embedded directly in the reconstruction pipeline. The neural network model enhances image quality at the early reconstruction level by removing image noise and ringing artifacts directly from the raw image data, providing consistently good, diagnostic-quality images even for challenging patients and difficult-to-image anatomies. The increase in SNR and image sharpness provided by AIR Recon DL also reduces scan times, which is very important for HN cancer patients.

 

Our HN acquisition protocol uses a 21-element head and neck coil configuration with a neurovascular adapter (Table 1, 2).

 

A table that has a number of different items on it

Table 1

 

A table that shows the differences between the two main types of a computer

Table 2

 

Diffusion-weighted imaging (DWI), based on echo-planar sequences, can be technically challenging due to its high sensitivity to noise and geometric distortion, which are consequences of susceptibility effects that disrupt magnetic field homogeneity. In HN MRI, these distortions are often caused by the numerous air-tissue interfaces linked to the complexity of the anatomical structure and the restorative materials used for dental treatments. Reduced field-of-view (FOV) DWI sequences, such as FOCUS, can improve imaging of the salivary glands and small lesions, reducing anatomical distortions. AIR Recon DL, compatible with FOCUS DWI sequences, can be activated in the reconstruction pipeline to minimize image noise.

Gallery Image 1
A
Gallery Image 2
B
Gallery Image 3
C

Figure 1.

These images demonstrate the value of imaging with T2 PROPELLER using AIR Recon DL in the evaluation of tumors in the (A) left hypopharynx with pathological neck nodes, (B) right larynx on the glottic region and (C) nasal cavity.

An ongoing study at our Institute evaluated the impact of AIR Recon DL on the ADC map

used by radiologists for differential diagnosis of lesions. Using a calibrated phantom (Diffusion Phantom, Caliber MRI, Boulder, CO), the study found minimal deviation in ADC values with and without DL application, with the most significant variations observed in higher noise DWI images. The preliminary findings suggested a tendency to a better agreement between nominal and measured values when using AIR Recon DL, suggesting that AIR Recon DL may improve SNR and enhance lesion margin visualization, leading to more accurate region-of-interest delineation (Figure 4)

Gallery Image 1
A
Gallery Image 2
B
Gallery Image 3
C
Gallery Image 4
D

Figure 2.

MR images of the base of the tongue in a patient with HPV-positive squamous cell carcinoma (SCC cT4a N0). Post processing was performed on GenIQ. (A) T2-weighted PROPELLER sequences with AIR Recon DL show an infiltration of the extrinsic muscles with the involvement of the lateral and posterior oropharynx wall. (B) T1-weighted FSPGR post-contrast shows the low enhancement of the tumor. (C) MUSE DWI shows a restricted signal intensity. (D) IAUGC (initial area under the gadolinium curve) map derived from DCE-MRI shows a hypovascular lesion.

Clinical and technical description

Advanced MR techniques, such as DWI and dynamic contrast-enhanced (DCE) MR, provide quantitative functional parameters such as tissue cellularity, vascular permeability/perfusion and hypoxia.1-3 These functional parameters are known to have radiobiological relevance and could potentially guide treatment adaptation based on baseline values and early changes during radiotherapy. DCE-MR was specifically designed to identify tumor vascular heterogeneity and in particular poorly oxygenated sub-volumes of the lesion. DCE-MR is the most frequently used perfusion technique for the HN region. This technique allows for the calculation of semiquantitative and quantitative parameters that have the potential to become a powerful diagnostic and therapeutic oncologic biomarker. It is also used for target delineation, treatment response assessment, and normal tissue injury evaluation.

 

In an ongoing research protocol at our Institute, we hypothesize that once the hypoxic sub-volumes within the primary tumor site are identified and localized, an intensified Intensity Modulated Radiation Thereapy (IMRT) regimen delivered to these regions will provide superior local control rates than the standard IMRT approach. Based on a pre-treatment DCE-MR study, patients with a hypoxic volume of at least 5 cc at the primary site will be considered eligible and randomized into either the standard or intensified IMRT regimens.

Gallery Image 1
A
Gallery Image 2
B
Gallery Image 3
C
Gallery Image 4
D

Figure 3.

MR T2-weighted PROPELLER with AIR Recon DL before radiotherapy shows (A) a large lesion of the right base of the tongue in a patient with HPV-positive SCC cT4. (B) The Ktrans map derived from DCE-MRI shows the presence of a hypovascular area into the tumor. (C) T2-weighted PROPELLER with the use of AIR Recon DL after 10 fractions shows a slight reduction in the tumor volume and (D) a more evident decrease of the hypo-oxygenated regions based on the Ktrans map (- 73.6%), with an increase of the entire tumor vascularization (+ 60.8%). Post processing was performed on GenIQ.

Improvement in perfusion parameters quantification can be achieved by calculating a T1 map at baseline using T1-weighted sequences with multiple flip-angles (FA); our protocol employs FA at 2° – 5° – 7° – 12° – 15° – 25° – 30° with a slice thickness of 4 mm and an acquisition time of 5 seconds per flip angle. DCE-MR is based on a 3D axial FSPGR with a temporal resolution of 5 seconds, a slice thickness of 4 mm, 20 slices per volume, 65 phases total, and an acquisition time of 5:24 minutes. We have used a clustering approach for the perfusion maps to evaluate the hypoperfused tumor sub-volume on the images. Clustering, also defined as habitat imaging techniques, has recently been proposed to assess intra-tumoral heterogeneity.4 Machine learning algorithms drive this emergent image quantitative analysis to recognize separated tumor components with similar characteristics.5 These tumor sub-volumes may identify tumor subareas with distinct biological and molecular profiles, suggesting that habit imaging techniques may be central in explaining the different outcomes of patients receiving the same treatment protocol.

Gallery Image 1
A
Gallery Image 2
B

Figure 4.

Contour delineation of the largest section of the lesion on the ADC map in diffusion-weighted images obtained with FOCUS and reconstructed (A) without and (B) with AIR Recon DL.

Advancements in imaging techniques at 3.0T also allow the acquisition of dedicated MR sequences to obtain a detailed neurography study. Optimizing these sequences is challenging because they need a large FOV, thin slices, high SNR and uniform fat and venous suppression.6 In oncologic applications, an MR neurography (MRN) study may support the surgeon in planning the surgical intervention, providing detailed information about the nerve location and possible involvement by the disease; it may also help the radiation oncologist to delineate sensible organs at risk, such as the brachial plexus, to reduce unnecessary organ irradiation if in proximity to the target volumes.

Gallery Image 1
A
Gallery Image 2
B
Gallery Image 3
C
Gallery Image 4
D

Figure 5.

MR Neurography (MRN). (A, B) MRN shows a schwannoma of the brachial plexus. (C, D) MRN shows the relationship between neck nodes and the brachial plexus.

What makes it innovative?

PROPELLER sequences are key to imaging anatomies susceptible to motion during MR examinations, particularly in subjects who struggle to remain still, such as those with HN tumors. Now with AIR Recon DL extended to the PROPELLER sequence, we can obtain motion-insensitive images with the same sharpness and diagnostic value as standard FSE, without worrying about repeating scans degraded by motion artifacts.

 

Perfusion MR allows for the assessment of tumor vascularization and has the potential to enhance the accuracy of MR in both diagnostic and prognostic evaluations; however, its role in HN tumors remains to be defined. Perfusion DCE may be recommended for diagnosis, radiotherapy planning, response assessment after chemo-radiotherapy or induction chemotherapy, and for detecting post-treatment local recurrence during follow-up. For these reasons, we encourage the adoption of a standardized use of perfusion-weighted imaging across research and clinical centers to produce more robust evidence, preferably through multicenter studies, of the clinical value of this challenging yet promising functional MR technique for future applications in HN tumors. 

 

 

 

 

A purple circle with the letter s in it

 

  1. Smits HJG, Vink SJ, de Ridder M, Philippens MEP, Dankbaar JW. Prognostic value of pretreatment radiological MRI variables and dynamic contrast-enhanced MRI on radiotherapy treatment outcome in laryngeal and hypopharyngeal tumors Clin Transl Radiat Oncol. 2024 Sep 12;49:100857.
  2. Hernandez-Herrera GA, Calcano GA, Nagelschneider AA, Routman DM, Van Abel KM. Imaging Modalities for Head and Neck Cancer: Present and Future Surg Oncol Clin N Am. 2024 Oct;33(4):617-649.
  3. Bane O, Hectors SJ, Wagner M, et al. Accuracy, repeatability, and interplatform reproducibility of T1 quantification methods used for DCE-MRI: Results from a multicenter phantom study. Magn Reson Med. 2018 May;79(5):2564-2575.
  4. Li S, Dai Y, Chen J, Yan F, Yang Y. MRI-based habitat imaging in cancer treatment: current technology, applications, and challenges. Cancer Imaging. 2024 Aug 15;24(1):107.
  5. van der Hulst HJ, Jansen RW, Vens C, et al. The Prediction of Biological Features Using Magnetic Resonance Imaging in Head and Neck Squamous Cell Carcinoma: A Systematic Review and Meta-Analysis. Cancers (Basel). 2023 Oct 20;15(20):5077.
  6. Van der Cruyssen F, Croonenborghs TM, Renton T, et al. Magnetic resonance neurography of the head and neck: state of the art, anatomy, pathology and future perspectives. Br J Radiol. 2021 Mar 1;94(1119):20200798.

 

Continue Reading
The power to explore further

The power to explore further

Volume 38 — Spring 2025

Published

AI synergistically transforms the MR experience

AI synergistically transforms the MR experience

Published

Join us in seeing the future of MR

Join us in seeing the future of MR

Published

How we are fueling further exploration for clinicians and scientists

How we are fueling further exploration for clinicians and scientists

Published

GE HealthCare welcomes new SIGNA Sprint 1.5T to explore further

GE HealthCare welcomes new SIGNA Sprint 1.5T to explore further

Published

In partnership with EU consortium, GE HealthCare to advance AI in healthcare through synthetic data

In partnership with EU consortium, GE HealthCare to advance AI in healthcare through synthetic data

Published

PREDICTOM launches prospective study, gears up for MR protocol harmonization

PREDICTOM launches prospective study, gears up for MR protocol harmonization

Published

First modular MR in Southeast Asia

First modular MR in Southeast Asia

Published

ISMRM 2025 abstracts

ISMRM 2025 abstracts

Published

MR neurography imaging for all: providing a new service line using existing sequences and systems

MR neurography imaging for all: providing a new service line using existing sequences and systems

Published

Progardia lowers energy consumption, reduces carbon footprint and increases image quality with continued investment in MR and AI

Progardia lowers energy consumption, reduces carbon footprint and increases image quality with continued investment in MR and AI

Published

Life-speed imaging with Sonic DL Cine opens up cardiac MR access to those who need it most

Life-speed imaging with Sonic DL Cine opens up cardiac MR access to those who need it most

Published

Elevating neuro imaging with an upgrade to SIGNA PET/MR

Elevating neuro imaging with an upgrade to SIGNA PET/MR

By Ju-Chieh Kevin Cheng, PhD, UBC PET/MRI Physicist, Research Scientist at UBC Pacific Parkinson’s Research Centre, Vesna Sossi, PhD, Professor, Department of Physics and Astronomy, Adjunct Professor at Djavad Mowafaghian Centre for Brain Health, and Elham Shahinfard, PhD, PET-MR Imaging Program Manager, Charles E. Fipke Integrated Neuroimaging Suite, The University of British Columbia, Vancouver, British Columbia, Canada

Published

Advancing women’s imaging through collaboration

Advancing women’s imaging through collaboration

Published

Leading the way in breast and pelvic imaging with SIGNA Hero

Leading the way in breast and pelvic imaging with SIGNA Hero

By Shunsuke Matsumoto, MD, Radiology Manager, Keiyu Hospital, Yokohama, Japan

Published

Shorter scan times and higher quality imaging reduce the need for sedation in pediatric patients

Shorter scan times and higher quality imaging reduce the need for sedation in pediatric patients

By Pär-Arne Svensson, BSc, Research Radiographer MRI, and Liz Ivarsson, MD, Senior Consultant and Pediatric Radiologist, Queen Silvia Children’s Hospital, Gothenburg, Sweden

Published

Enhancing surgical precision and patient outcomes with fMRI

Enhancing surgical precision and patient outcomes with fMRI

By Yoshiyuki Watanabe, MD, PhD, Professor, Department of Radiology, Shiga University of Medical Science, Ōtsu, Shiga, Japan

Published

Bringing AIR Recon DL to multi-shot diffusion

Bringing AIR Recon DL to multi-shot diffusion

By Patricia Lan, PhD, MR Scientist, Arnaud Guidon, PhD, Director of Body & Oncology MR, and Suryanarayanan “Shiv” Kauskik, PhD, Senior Digital Product Manager, GE HealthCare

Published

Ultra-high contrast MR of the brain and spinal cord

Ultra-high contrast MR of the brain and spinal cord

By Paul Condron, BSc (Hons), MRI Charge Technologist, Mark Bydder, PhD, Senior Research Fellow, Samantha J. Holdsworth, PhD, Executive Director at Mātai Medical Research Institute and Associate Professor at the University of Auckland, Daniel M. Cornfeld, MD, FRANZCR, Clinical Lead, and Graeme M. Bydder, FRANZCR, Scientific Advisory Board Member, Mātai Medical Research Institute, Gisborne, New Zealand

Published

Deuterium metabolic imaging shows promise in detecting Alzheimer’s disease

Deuterium metabolic imaging shows promise in detecting Alzheimer’s disease

Published