Cancer patients are often referred for whole-body imaging to evaluate the presence of secondary or metastatic lesions, which can change the course of patient treatment. In our institution, whole-body diffusion MR is our preferred modality for this type of study.
Diffusion-weighted imaging (DWI) on MR has shown excellent sensitivity (82%) and specificity (97%) compared to PET/CT (sensitivity of 72% and specificity of 92%) in whole-body imaging in melanoma metastases.1 Additionally, DWI MR has been shown to be the most accurate for detecting metastases in the bone, liver, subcutaneous and intra-peritoneal sites compared to PET/CT.1
However, non-optimal DWI MR acquisitions can produce scintigraphy-like images that may produce false positive results. Combining DWI MR with an inversion time technique can help overcome this issue by allowing for a better suppression of fat signal, which helps suppress background body signal. Further, utilizing an MR system that has strong magnet homogeneity allows for the acquisition of large field-of-view (FOV) images and helps reduce distortions in areas of off-center imaging, such as the extremities. DWI STIR with a "3-in-1" technique provides an excellent compromise between acquisition time and signal-to-noise ratio (SNR), allowing acquisition of thin slices that may help further enhance sensitivity of the DWI sequence.
Patient history
A 75-year-old female with known melanoma was referred for whole-body DWI MR for evaluation of suspected metastatic (secondary) lesions on her extremities. Post-contrast whole-body MR with LAVA Flex was also used to evaluate areas with suspicious contrast uptake.
MR findings
- Important effusion on the glenohumeral joint of the right shoulder, to be investigated
- Several bilateral swollen axillary lymph nodes, predominant in number and size at the right axillary region
- Lesion located on the proximal metaphysis of the ulna of the left arm and suspected to be metastatic, measuring 13 mm from its long axis, hyperintense in diffusion, hyperintense in STIR and gadolinium-enhanced on post-contrast T1 sequence
- Second suspicious lesion with similar aspect, measuring 9 mm and located on the distal diaphysis of the radius of the right arm
- Lesion on the left iliac joint, measuring 6 mm, hyperintense in diffusion, with no aspects on the other sequences
- Small nodule of 15 mm situated on the soft tissues of the right thigh, at the inferior part located within the posterior muscular compartment, hyperintense in diffusion.





Figure 1.
Whole-body MR imaging with approximately 120 cm coverage. (A-B) Coronal in-phase and water LAVA Flex; 2.5 x 3 x 2.6 mm, 3 stations, 1:45 min.; (C) Coronal STIR, 1.5 x 2.1 x 8 mm, 3 stations, 5:30 min.; and (D) Coronal water LAVA Flex, post contrast, 2.5 x 3 x 2.6 mm, 3 stations, 1:45 min.



Figure 2.
Whole-body MR imaging with approximately 120 cm coverage. Binded Axial diffusion STIR 3-in-1, b600, 4.4 x 4.4 x 3 mm, 20 stations, 36:20 min. (A-B) MIP projections; and (C) volume rendered image. Red circle depicts the effusion on the glenohumeral joint of the right shoulder. Red arrows show the bilateral swollen axillary lymph nodes. Blue circle denotes the lesion on the left iliac joint.






Figure 3.
(A) Coronal STIR, 1.5 x 2.1 x 8 mm; (B) same image with color map; (C-D) diffusion STIR 3-in-1 b600, 4.4 x 4.4 x 3 mm; and (E-F) Coronal water LAVA Flex post contrast, 2.5 x 3 x 2.6 mm and Axial MPR. Lesion located on the proximal metaphysis of the ulna of the left arm (red circle), suspected to be metastatic.


Figure 4.
(A) Fusion STIR + diffusion STIR 3-in-1 b600; (B) diffusion STIR b600, 4.4 x 4.4 x 3 mm. Small nodule was located on the soft tissues of the right thigh (red circle).
Discussion
The patient has several small lesions suspected to be secondary bone lesions from her primary cancer, melanoma. Having the ability to detect these lesions on the patient’s arms, iliac joint and thigh will enable the oncologist to re-stage the cancer and adapt therapy.
Whole-body DWI is an excellent option for evaluation of metastatic bone lesions compared to PET/CT and scintigraphy. It provides good image quality and provides the information clinicians need for patient management.
Even if PET/CT were recommended, this type of system is not available in all regions and is often more expensive than MR. Scintigraphy exposes the patient to additional radiation and MR has been found to provide similar sensitivity and specificity for patients with a small number of bone lesions, 58% sensitivity/33% specificity for whole-body DWI MR versus 67% sensitivity/58% specificity for scintigraphy; and patients with multiple bone lesions, 97% sensitivity/91% specificity for whole-body DWI MR versus 48% sensitivity/42% specificity for scintigraphy.2
With MR, we can avoid patient exposure to additional radiation dose and achieve a high sensitivity and specificity for detection of metastatic lesions. Additionally, the strong magnet homogeneity on the SIGNA™Explorer helps to decrease distortion when imaging large FOVs and "3-in-1" diffusion STIR provides the thin-slice data needed for a confident diagnosis in shorter scan times.

References
- Laurent V, Trausch G, Bruot O, Oliver P, Felblinger J, Regent D. Comparative study of two whole-body imaging techniques in the case of melanoma metastases: advantages of multi-contrast MRI examination including a diffusion-weighted sequence in comparison with PET-CT. Eur J Radiol. 2010 Sep;75(3):376-83.
- Skeletal Radiol. 2010 Apr;39(4):333-43. doi: 10.1007/s00256-009-0789-4. Gutzeit A, Doert A, Froehlich JM, et al. Comparison of diffusion-weighted whole-body MRI and skeletal scintigraphy for the detection of bone metastases in patients with prostate or breast carcinoma. Skeletal Radiol. 2010 Apr;39(4):333-43.

Pulse of MR
Volume 26 – Spring 2019
Published
MR is elevating radiology and patient care
Published

Artful images
Published

Global MR market size to grow
Published

Leading the AI charge
Published

More comfortable prostate scans
Published

Clear the AIR
Published

EPI in the mix
Published

Video game tech fueled ViosWorks
Published

European team awarded 3.2M grant
Published

MR optimizes diagnoses and long-term patient management
Published

Localizing pain with PET/MR
Published

New software release focuses on enhancements
Published

Development of an MR-only radiation therapy workflow
-
Published

Thirty-minute PET/MR exam for pediatric cancer patients
By Jing Qi, MD, Assistant Professor, and Nghia (Jack) Vo, MD, Chief of Pediatric Radiology, Medical College of Wisconsin
Published

Transforming the MR imaging experience for one of Sweden’s largest pediatric hospitals
-
Published

AIR Technology: a brilliant improvement in high-quality imaging and patient comfort
-
Published

New deep learning tool streamlines MR slice prescription
Published

An upgrade that meets the expectation for higher resolution, SNR and productivity
-
Published

Ultra-flexible AIR Technology Suite making a difference in the technologist’s workflow
-
Published

MR Excellence Program is a beacon for optimizing MR imaging workflow and the patient experience
-
Published

Cardiac MR in patients with CIEDs
By Karl Vigen, PhD, Senior Scientist, and Christopher Francois, MD, Professor, University of Wisconsin Hospital, Madison, WI
Published

Body imaging with AIR Technology Anterior Array and Posterior Array
Submitted by Quirónsalud Madrid University Hospital
Published

Neuro imaging with 48-channel Head Coil
By Krisztina Baráth, MD, neuroradiologist, and Brigitte Trudel, RT(R)(MR), MRI Chief Technologist, RNR Institute of Radiology and Neuroradiology at Glattzentrum
Published

Diffusion imaging with AIR Technology Suite
Submitted by Kawasaki Saiwai Hospital
Published

Free-breathing liver imaging using DISCO with Auto Navigator
By Claire Moisson, RT(R)(MR), and Stephanie Sellier, RT(R)(MR), Lead Technologist, Medipole de Savoie
Published

Free-breathing navigator-echo triggered diffusion-weighted imaging in the evaluation of hepatobiliary disease
By David Bowden FRCR, Consultant Hepatobiliary & GI Radiologist, Addenbrooke’s Hospital, Cambridge University Hospitals NHS Foundation Trust
Published

3D MRCP with HyperSense: an evaluation of respiratory-triggered and breath-hold sequences
By Marc Zins, MD, Head of the Radiology Department, Saint-Joseph Hospital
Published

Detecting ischemia-induced cardiac fibrosis with phase sensitive MDE
By Abdelhamid Derriche, MD, site radiologist, and Orkia Ferdagha, MR technologist, PRIISM, EHP Kara
Published

SIGNA™ Works: tuned for productivity and efficient workflows
By Steve Lawson, RT(R)(MR), Global MR Clinical Marketing Manager, and Heide Harris, RT(R)(MR), Global Product Marketing Director, MR Applications and Visualization, GE Healthcare
Published

Hyperpolarized gas lung imaging
-
Published